The Performance of Fine Resolution Neutron Powder Diffractometer at PTBIN-BATAN
Main Authors: | Fajar, A.; Center for Technology of Nuclear Industry Materials, BATAN Puspiptek Serpong, Tangerang, Indonesia, Mugirahardjo, H.; Center for Technology of Nuclear Industry Materials, BATAN Puspiptek Serpong, Tangerang, Indonesia |
---|---|
Other Authors: | National Nuclear Energy Agency of Indonesia |
Format: | Article info application/pdf eJournal |
Bahasa: | eng |
Terbitan: |
PPIKSN-BATAN
, 2010
|
Subjects: | |
Online Access: |
http://aij.batan.go.id/index.php/aij/article/view/5 http://aij.batan.go.id/index.php/aij/article/view/5/1 |
Daftar Isi:
- We describe the performance of a fine resolution neutron powder diffractometer, DN3, which has been installed at the neutron guide (NG2) in the Neutron Guide Hall about 71 m away from the Multipurpose G. A. Siwabessy reactor core. DN3 has a multi-detector system which consists of 32 3He neutron detectors and soller-type collimation. The resolution curve of the instrument was found to have little variation over a wide angular region. The best resolution is Δ2θ (FWHM) of 0.23o and Δd / d of 5x10-3, where 2θ and d are scattering angle and d-spacing, respectively. The instrument provides sample environment of low temperatures using a cryostat having a temperature range of 10-300K. Several experimental results, interlaboratory comparison study and he next activities to improve the performance of DN3 equipment have also been discussed. Received: 15 October 2009; Revised: 12 April 2010; Accepted: 21 April 2010