The Optimization of Collimator Material and In Vivo Testing Dosimetry of Boron Neutron Capture Therapy (BNCT) on Radial Piercing Beam Port Kartini Nuclear Reactor by Monte Carlo N-Particle Extended (MCNPX) Simulation Method

Main Authors: Sardjono, Yohannes, Kusminarto, Kusminarto, Wusko, Ikna Urwatul
Format: Article info application/pdf eJournal
Bahasa: eng
Terbitan: Fakultas Sains dan Matematika Universitas Kristen Satya Wacana , 2018
Online Access: https://ejournal.uksw.edu/ijpna/article/view/1673
https://ejournal.uksw.edu/ijpna/article/view/1673/894
Daftar Isi:
  • Boron Neutron Capture Therapy (BNCT) on radial piercing beam port Kartini nuclear reactor by MCNPX simulation method has been done in the National Nuclear Energy Agency Yogyakarta. BNCT is a type of therapy alternative that uses nuclear reaction 10B (n, α) 7Li to produce 2.79 MeV total kinetic energy. To be eligible IAEA conducted a study of design improvements and variations on some parameters to optimum condition which are Ni-nat thickness of 1.75 cm as collimator wall, Al2S3 as thick as 29 cm as moderator, Al2O3 0.5 cm thick as filter, Pb and Bi thickness of 4 cm as the end of the gamma shield collimators and Bi thickness of 1.5 cm as the base gamma shield collimators. The total dose was accepted in the tumor tissue 900 × 10-4 Gy/s. Radiation dose on the tumor tissue is 50±3 Gy with time irradiation of 9 minutes and 10 seconds. That dose was given into skin tissue and healthy liver tissue consecutively (6.00±0.05) × 10-2 Gy and (10.00±0.05) × 10-2 Gy. It shows the dose received by healthy tissue is still within safe limits.