APPLICATION OF CENTRAL COMPOSITE DESIGN FOR OPTIMIZATION THE DETERMINATION OF LEAD USING ADSORPTIVE CATHODIC STRIPPING VOLTAMMETRY

Main Authors: Deswati, Deswati, Rahmi HG, Izzati, Suyani, Hamzar, Zein, Rahmiana, Alif, Admin
Format: Article PeerReviewed Book
Bahasa: eng
Terbitan: Rasayan Journal , 2016
Subjects:
Online Access: http://repo.unand.ac.id/5230/1/Rasayan%20_Vol.9%2C%20No.1%2C%208-17%2C%20Jan.-March%2C%202016%2C%20RJC-1380.pdf
http://repo.unand.ac.id/5230/
http://www.rasayanjournal.com
Daftar Isi:
  • This paper describes an optimization method for rapid determination of lead from Maninjau lake waters by adsorptive cathodic stripping voltammetry using central composite design. The influence of several parameters were studied : variations of calcon concentration, pH, accumulation potential and accumulation time. The design study was a central composite design with 4 factors/variables 3 levels and 31 treatment combinations. From analysis of variance, concluded to accept the second-order model and a significant effect on the response variable (peak current). Based on central composite design, obtained the optimum conditions of lead were : concentration of calcon 0.76 mM, pH = 5.88, accumulation potential -0.45 Volt and accumulation time 88.38 seconds with a maximum peak current 42.56 nA. This method has been successfully applied to water was obtained : relative standard deviation 1.5%, recovery 97.33%, the linear range 0.2-105 μ g/L, coefficient of determination 0.92 with a detection limit 1.02 μ g/L. Keywords: central composite design, lead, adsorptive cathodic stripping voltammetry