VALIDASI METODE KALIBRASI GAS ANALYZER UNTUK PENGUKURAN O2, CO, NO, NO2, SO2, CH4, DAN H2S SECARA PERBANDINGAN LANGSUNG DENGAN CERITIFIED SPAN GAS

Main Authors: Hadi, Anwar, Subarja, Jaja Ahmad, Firdaus, Idris
Format: Article info application/pdf eJournal
Bahasa: ind
Terbitan: Pusat Penelitian dan Pengembangan Kualitas dan Laboratorium Lingkungan (P3KLL) , 2018
Subjects:
Online Access: http://ejournal.forda-mof.org/ejournal-litbang/index.php/JKLH/article/view/4890
http://ejournal.forda-mof.org/ejournal-litbang/index.php/JKLH/article/view/4890/4358
ctrlnum --ejournal.forda-mof.org-ejournal-litbang-index.php-index-oai?verb=ListSets:article-4890
fullrecord <?xml version="1.0"?> <dc schemaLocation="http://www.openarchives.org/OAI/2.0/oai_dc/ http://www.openarchives.org/OAI/2.0/oai_dc.xsd"><title lang="id-ID">VALIDASI METODE KALIBRASI GAS ANALYZER UNTUK PENGUKURAN O2, CO, NO, NO2, SO2, CH4, DAN H2S SECARA PERBANDINGAN LANGSUNG DENGAN CERITIFIED SPAN GAS</title><creator>Hadi, Anwar</creator><creator>Subarja, Jaja Ahmad</creator><creator>Firdaus, Idris</creator><subject lang="id-ID">method validation; certified span gas; zero gas; gas analyzer; calibration and measurement capability</subject><description lang="id-ID">Salah satu persyaratan validasi data pengukuran gas adalah dilakukan oleh personel yang kompeten dengan menggunakan gas analyzer yang terkalibrasi dan pengunaan metode tervalidasi. Validasi metode pengukuran gas dengan peralatan gas analyzer dilakukan secara perbandingan langsung dengan certified span gas (CSG) yang tertelusur ke sistem satuan internasional melalui rantai perbandingan yang tidak terputus. Tujuan validasi metode kalibrasi adalah untuk mendapatkan informasi penting dalam menilai kemampuan serta keterbatasan metode kalibrasi. Validasi metode kalibrasi gas analyzer O2, CO, NO, NO2, SO2, CH4, dan H2S menghasilkan akurasi, presisi dan linieritas memenuhi kriteria batas keberterimaan yang telah ditentukan yaitu % R &#x2264; 2% relatif, % RSD &#x2264; 0,5 nilai horwitz dan koefisien determinasi &#x2265; 0,990. Kemampuan kalibrasi dan pengukuran digunakan sebagai informasi atas kemampuan pengukuran dan kalibrasi serta ketidakpastian yang dicapai yang tersedia bagi pelanggan dalam kondisi normal. Dengan demikian, metode kalibrasi gas analyzer yang telah divalidasi dapat digunakan untuk mengukur sampel gas untuk kebutuhan pelanggan</description><publisher lang="en-US">Pusat Penelitian dan Pengembangan Kualitas dan Laboratorium Lingkungan (P3KLL)</publisher><contributor lang="id-ID"/><date>2018-10-01</date><type>Journal:Article</type><type>Other:info:eu-repo/semantics/publishedVersion</type><type>Other:</type><type>File:application/pdf</type><identifier>http://ejournal.forda-mof.org/ejournal-litbang/index.php/JKLH/article/view/4890</identifier><identifier>10.20886/jklh.2017.11.2.62-71</identifier><source lang="en-US">Jurnal Ecolab; Vol 11, No 2 (2017): Jurnal Ecolab; 62-71</source><source lang="id-ID">Jurnal Ecolab; Vol 11, No 2 (2017): Jurnal Ecolab; 62-71</source><source>2502-8812</source><source>1978-5860</source><language>ind</language><relation>http://ejournal.forda-mof.org/ejournal-litbang/index.php/JKLH/article/view/4890/4358</relation><rights lang="en-US">Copyright (c) 2018 Jurnal Ecolab</rights><rights lang="en-US">http://creativecommons.org/licenses/by-nc-sa/4.0</rights><recordID>--ejournal.forda-mof.org-ejournal-litbang-index.php-index-oai?verb=ListSets:article-4890</recordID></dc>
language ind
format Journal:Article
Journal
Other:info:eu-repo/semantics/publishedVersion
Other
Other:
File:application/pdf
File
Journal:eJournal
author Hadi, Anwar
Subarja, Jaja Ahmad
Firdaus, Idris
title VALIDASI METODE KALIBRASI GAS ANALYZER UNTUK PENGUKURAN O2, CO, NO, NO2, SO2, CH4, DAN H2S SECARA PERBANDINGAN LANGSUNG DENGAN CERITIFIED SPAN GAS
publisher Pusat Penelitian dan Pengembangan Kualitas dan Laboratorium Lingkungan (P3KLL)
publishDate 2018
topic method validation
certified span gas
zero gas
gas analyzer
calibration and measurement capability
url http://ejournal.forda-mof.org/ejournal-litbang/index.php/JKLH/article/view/4890
http://ejournal.forda-mof.org/ejournal-litbang/index.php/JKLH/article/view/4890/4358
contents Salah satu persyaratan validasi data pengukuran gas adalah dilakukan oleh personel yang kompeten dengan menggunakan gas analyzer yang terkalibrasi dan pengunaan metode tervalidasi. Validasi metode pengukuran gas dengan peralatan gas analyzer dilakukan secara perbandingan langsung dengan certified span gas (CSG) yang tertelusur ke sistem satuan internasional melalui rantai perbandingan yang tidak terputus. Tujuan validasi metode kalibrasi adalah untuk mendapatkan informasi penting dalam menilai kemampuan serta keterbatasan metode kalibrasi. Validasi metode kalibrasi gas analyzer O2, CO, NO, NO2, SO2, CH4, dan H2S menghasilkan akurasi, presisi dan linieritas memenuhi kriteria batas keberterimaan yang telah ditentukan yaitu % R ≤ 2% relatif, % RSD ≤ 0,5 nilai horwitz dan koefisien determinasi ≥ 0,990. Kemampuan kalibrasi dan pengukuran digunakan sebagai informasi atas kemampuan pengukuran dan kalibrasi serta ketidakpastian yang dicapai yang tersedia bagi pelanggan dalam kondisi normal. Dengan demikian, metode kalibrasi gas analyzer yang telah divalidasi dapat digunakan untuk mengukur sampel gas untuk kebutuhan pelanggan
id IOS3429.--ejournal.forda-mof.org-ejournal-litbang-index.php-index-oai?verb=ListSets:article-4890
institution Kementerian Lingkungan Hidup dan Kehutanan Republik Indonesia
institution_id 475
institution_type library:special
library
library Perpustakaan Kementerian Lingkungan Hidup dan Kehutanan RI
library_id 121
collection Indonesian Journal of Forestry Research
repository_id 3429
subject_area Kehutanan
Lingkungan
Pertanian
city BOGOR
province JAWA BARAT
repoId IOS3429
first_indexed 2019-05-06T00:36:20Z
last_indexed 2019-05-06T00:36:20Z
recordtype dc
_version_ 1764501244702883840
score 17.538404