Study on Humidity Inhomogeneity Measurement of a Climatic Chamber

Main Author: Tistomo, Arfan Sindhu; Puslit KIM LIPI, Kompleks Puspiptek Serpong, Tangerang
Format: Article info application/pdf Journal
Bahasa: eng
Terbitan: LIPI Press, Anggota IKAPI , 2013
Online Access: http://jurnalinstrumentasi.bsn.go.id/index.php/ji/article/view/37
http://jurnalinstrumentasi.bsn.go.id/index.php/ji/article/view/37/16
Daftar Isi:
  • Study on a method for humidity inhomogeneity measurement based on temperature inhomogeneity has been done. The method combines AS 2853 and DKD 2009 guidelines. A manufacturer specification of a climatic chamber is used for the reference. A Monte Carlo simulation technique is used to generate random data having rectangular distribution. As a result, the introduced method is able to describe the climatic chamber specification. In addition, sensitivity coefficient of relative humidity to temperature varies from 0.02%RH/°C to 7%RH/°C for temperature of 10°C~100°C with relative humidity of 20%RH~98%RH. The worst humidity inhomogeneity for a climatic chamber can be found at lower limit temperature coupled with upper limit relative humidity.